Randox Laboratories Ltd. recalls Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567

Recalled April 24, 2024 · FDA recall Z-1593-2024 · Randox Laboratories

Hazard
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%. · FDA Class II
Units
8 units
Sold at
US Nationwide distribution in the states of GA, MA, WV including PR.

Description

Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.

Product
Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
Manufacturer(s)
Randox Laboratories Ltd.
Made in
Ireland