RecallWatchUSA

Home / Recalls / Medical Devices

Randox Laboratories Ltd. recalls Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena

Recalled August 16, 2023 · FDA recall Z-2362-2023 · Randox Laboratories

Hazard
Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results. · FDA Class II
Units
37 US 194 PR
Sold at
US Nationwide distribution in the states of CA, DC, GA, OH, OK, MN, MO, NJ, VA and Puerto Rico.

Description

Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results.

Product
Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)
Manufacturer(s)
Randox Laboratories Ltd.
Made in
Ireland

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Randox Laboratories Ltd. recalls Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena | RecallWatchUSA