Randox Laboratories Ltd. recalls Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 c
Recalled May 15, 2024 · FDA recall Z-1759-2024 · Randox Laboratories
Hazard
Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods. · FDA Class II
Units
80,366 units (US: 242 units; OUS: 80,094 units)
Sold at
Domestic: Nationwide Distribution.
Description
Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.
Product
Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinical chemistry tests.
Manufacturer(s)
Randox Laboratories Ltd.
Made in
Ireland