Steris Recalls

113 recalls on record · August 8, 2012 to June 24, 2026 · Drugs, Medical Devices

Recalls by year

12
13
14
15
16
18
19
20
21
22
23
24
25
26

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

May 6, 2015Medical DevicesFDA
Steris Corporation recalls AMSCO C and AMSCO 400 Steam Sterilizers, STERIS Corporation. AMSCO C Small Sterilizer, AMSCO 400 Small Sterilizer, AM
Hazard: STERIS has identified that the control board software in select AMSCO 400 and AMSCO C units will interrupt and cancel a processing cycle should the selected sterilizer shutdown time coincide with a processing cycle. This may result in a procedure delay.
May 6, 2015Medical DevicesFDA
Steris Corporation recalls Caviwave Pro Ultrasonic Cleaning System, Product Usage: To wash (remove tissue, blood and other contaminants) or dr
Hazard: Should the Caviwave Pro Ultrasonic Cleaning System experience a temporary interruption in electrical power during a processing cycle, the circulation pump may stop functioning. In addition, it is possible that the ultrasonic generator may stop working. The machine does not provide an alert notifying the user of this device malfunction.
January 14, 2015Medical DevicesFDA
Steris Corporation recalls Harmony¿ vLED Surgical Lighting System, one unit packaged per box The Harmony¿ vLED Surgical Lighting System is a varia
Hazard: STERIS has learned that groups of LED lights (modules) contained within the vLED light head may illuminate inconsistently.
November 19, 2014Medical DevicesFDA
Steris Corporation recalls The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal tab
Hazard: The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.
July 16, 2014Medical DevicesFDA
Steris Corporation recalls Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/D
Hazard: STERIS has identified that deformation of the washers door handle pin may occur when the door handle is rotated with sufficient force. Should deformation and/or breakage of the door handle pin occur there is a potential for the washer door to become inoperable.
July 16, 2014Medical DevicesFDA
Steris Corporation recalls The STERIS 4085 General Surgical Table Product Usage: is an electro-hydraulically operated surgical table designed t
Hazard: STERIS engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked.
July 2, 2014Medical DevicesFDA
Steris Corporation recalls Vertical Spring Arm for Single or Dual Monitor Mount (Harmony LL 500 and Harmony LL 700 surgical lighting systems), Ster
Hazard: Potential for the welded seam in affected monitor spring arms to crack and/or break.
March 19, 2014Medical DevicesFDA
Steris Corporation recalls VERIFY¿ SixCess 270FP Challenge Pack, STERIS Corporation, 5960 Heisley Road, Mentor, OH 44060, 440-354-2600. The VERI
Hazard: STERIS has identified that the outer shipping package for one lot (Lot #024552A) of VERIFY¿ SixCess Challenge Packs incorrectly identifies the lot as Lot #024522A.
January 8, 2014Medical DevicesFDA
Steris Corporation recalls SYSTEM 1E Liquid Chemical Sterilant Processing System, Distributed by STERIS Corporation, Mentor, OH The SYSTEM 1E Li
Hazard: Customers indicated fluctuations in environmental conditions sites and gradual degradation/deformation of check valve o-rings following frequent contact with S40 sterilant use solution result in nuisance aborts.
January 8, 2014Medical DevicesFDA
Steris Corporation recalls VERIFY¿ Dual Species Self-Contained Biological Indicator, Distributed by STERIS Corporation, Mentor, OH. Intended for
Hazard: STERIS has identified that the population of the G. stearothermophilus spores that monitor STEAM sterilization cycles does not meet specifications for Lot # 131004 of the Verify Dual Species Self-Contained Biological Indicators. As a result the spore count on the affected product is lower than the labeled minimum and the product may incorrectly indicate that a 5-log spore kill was achieved.
December 18, 2013Medical DevicesFDA
Steris Corporation recalls AMSCO¿ Evolution¿ steam sterilizer Transfer Carriage. Product Usage: Mobile, wheeled transportation unit utilized t
Hazard: The firm discovered that unlatching can occur when overloaded transfer carriages are pulled or maneuvered incorrectly or with excessive force.
November 13, 2013Medical DevicesFDA
Steris Corporation recalls Harmony Equipment Management System, Model #603X/80XX, STERIS Corporation, Mentor, OH Harmony Equipment Management Sy
Hazard: Dr¿ger Medical GmbH has notified STERIS of the potential for a break in the drive screw contained in the motorized ceiling supply unit, a component of the Harmony EMS system. The ceiling supply unit controls the motorized adjustment of the height and rotation of the supply column. Should a break occur in the drive screw, there is potential for the adjustable arm to drop down, causing the EMS sys
October 2, 2013Medical DevicesFDA
Steris Corporation recalls Reliance Endoscope Processing System (REPS), STERIS Corporation, 5960 Heisley Road, Mentor, OH 44060-1834 USA www.steris
Hazard: The air pipe that delivers unfiltered ambient air from the procedure room into the water bottle on the listed FUJINON G5 & G8 type flexible GI endoscopes does not connect to the main air/water channel within the light guide (LG) connection head and, as a result, the air pipe does not receive forced flow during the Reliance EPS processing cycle.
July 31, 2013Medical DevicesFDA
Steris Corporation recalls STERIS Fine Traction Device, one unit per box Product Usage: The Fine Traction Device is used to provide traction an
Hazard: A bearing used in the assembly of the Fine Traction Device did not perform successfully in pre-release reliability testing; a replacement bushing passed reliability testing and was substituted per a change order. The change order authorized use of the substitute bushing in assembly of the Fine Traction Accessory; however, inadvertently the new part was not implemented. During complaint evaluati
July 3, 2013Medical DevicesFDA
Steris Corporation recalls Quick Connect QLC1677E, QRC1699E, QPC1721E and QFC1683E. STERIS Corporation, Mentor, OH 44060 QLC1677E, QRC1699E, QPC
Hazard: During a routine label review, the firm identified four SYSTEM 1E Quick Connects that will be updated to reflect the removal of obsolete device models. This labeling review process also highlighted the opportunity to combine two existing quick connects into a single product while eliminating the obsolete models.
March 13, 2013Medical DevicesFDA
Steris Corporation recalls AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthc
Hazard: The control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks up and cannot be used until serviced, causing Customer nuisance.
September 26, 2012Medical DevicesFDA
Steris Corporation recalls STERIS 5085 and 5085 SRT Surgical Tables, Product Usage: The 5085 is a general surgical table with high patient w
Hazard: STERIS has learned from Customer feedback and field service experience that damage from external impact to the table hydraulic column can cause the stainless steel shroud sections covering the table hydraulic column to become misaligned; if the damage is severe, the up/down movement of the table may be impeded. In addition, the table hand control may indicate that the table floor locks are unlock
September 26, 2012Medical DevicesFDA
Steris Corporation recalls Verify¿ Bowie-Dick Test Card, Model #S3098, Steris Biological Operations, 9325 Pinecone Drive, Mentor, OH 44060 The V
Hazard: STERIS has learned that the Verify Bowie Dick Test Cards are not performing to product specifications. Specifically, test cards have resulted in false fail results when sterilizer performance is within acceptable ranges.
September 26, 2012Medical DevicesFDA
Steris Corporation recalls A mobile, electro-hydraulically operated, fluoroscopic/C-Arm compatible general surgical table. Product Usage: Pro
Hazard: Customers may be storing objects on the base and/or around the column of the surgical table. This prohibited practice and misuse known by the customers can damage the override switch assembly located at the top of the column cover. Depending upon the nature and severity of the damage to the electronics of the table, it is possible that the table could move without prompting by an operator .
September 19, 2012Medical DevicesFDA
Steris Corporation recalls The SYSTEM 1E Liquid Sterilant Processing System. The SYSTEM 1E Liquid Sterilant Processing System is intended for th
Hazard: The firm initiated a recall after complaint investigations revealed that when the units are left on and not in use at night and/or on weekends the hose disconnects/separates resulting in water leakage that can cause damage to property.
September 12, 2012Medical DevicesFDA
Steris Corporation recalls 100 ft roll of tubing inside a cardboard box. Medical devices are sealed into the tubes prior to ethylene oxide steriliz
Hazard: On 7/20/2012 the firm became aware that a lot of tubing was distributed by the supplier after it was rejected by Steris due to failing process indicators.
September 5, 2012Medical DevicesFDA
Steris Corporation recalls QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical
Hazard: STERIS has been actively seeking devices for quick connect revalidation projects. STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design of seven (7) models in one quick connect. Due to the unavailability of these two validation devices, STERIS will remove 7 models from the labeling of QKC1692E and will re-label all QKC1692Es in the field..
August 8, 2012Medical DevicesFDA
Steris Corporation recalls The Reliance 130 and 130L Cart Washer/Disinfector is a high capacity mechanical washer intended for use in the washing,
Hazard: To ensure that users are properly operating the Reliance 130 and 130L Cart Washers, a door close confirmation button and door open delay will be installed on your cart washer.