December 17, 2014Medical DevicesFDA
Stryker Endoscopy recalls 1488 HD 3-Chip Inline Camera; Model Number 1488-710-105 - HD 3-Chip Inline Camera, C-Mount; Product Usage: High de
Hazard: In certain operating room environments, a colored screen image occurs with the 1488 Cameras that were manufactured during a specific time period
December 17, 2014Medical DevicesFDA
Stryker Endoscopy recalls 1488 HD 3-Chip Camera; Model Number 1488-210-105 - C-Mount. Product Usage: High definition camera to view endosco
Hazard: In certain operating room environments, a colored screen image occurs with the 1488 Cameras that were manufactured during a specific time period
July 30, 2014Medical DevicesFDA
Stryker Endoscopy recalls Disposable StrykeFlow Tip, Stryker Disposable Suction/I irrigator Tip; Sterile, Rx only, Latex Free. Stryker Suction Irr
Hazard: Internal review found that six lots of Disposable StrykeFlow Tips were shipped with a potentially deformed blister pouch which could impact sterility of the product.
July 2, 2014Medical DevicesFDA
Stryker Endoscopy recalls Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shav
Hazard: STERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing.
May 14, 2014Medical DevicesFDA
Stryker Endoscopy recalls 12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instrum
Hazard: Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.
October 2, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Vision Mounting Arm, for use with other Stryker Endoscopy medical device products for human patients. Model num
Hazard: When positioning the display located on top of the video cart, it is possible for the user to apply excessive force and push the display up and out of the mounting arm. This can result in the display falling upon the user/patient.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 3mm X 32cm Regular Tip, Non-Vented, Metal Finish. Suction irrigator tips aid in cle
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 3mm X 32cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleanin
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip with No Holes, Non-Reflective Finish. Suction irrigator tips
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: Vented Tip, 5mm X 45cm Non-Reflective with 30¿ Downward Bend At Proximal End. Sucti
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 5mm X 45cm Regular Tip, vented, Non-Reflective Finish. Suction irrigator tips aid i
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 5mm X 32cm Hydrodissection Tip, Non-Reflective Finish. Suction irrigator tips aid i
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 10mm X 32cm Regular Tip, vented, Non-Reflective Finish. Suction irrigator tips aid
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip, vented, Non-Reflective Finish. Suction irrigator tips aid i
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 5mm X 28cm Regular Tip, vented, Non-Reflective Finish. Suction irrigator tips aid i
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 2mm X 34cm Regular Tip with No Hole, Metal Finish. Suction irrigator tips aid in cl
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 2mm X 32cm Regular Tip with 2 Hole, Metal Finish. Suction irrigator tips aid in cle
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 2mm X 32cm Regular Tip with 1 Hole, Metal Finish. Suction irrigator tips aid in cle
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 5mm X 32cm Aspiration Needle Tip. Suction irrigator tips aid in cleaning and cleari
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 5mm X 24cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleanin
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip , Non-Vented, Metal Finish. Suction irrigator tips aid in cl
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 5mm X 45cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleanin
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 5mm X 32cm Tip Hydrodissection Tip, Metal Finish. Suction irrigator tips aid in cle
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 10mm X 32cm Pool Suction Tip, Metal Finish. Suction irrigator tips aid in cleaning
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleanin
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
May 15, 2013Medical DevicesFDA
Stryker Endoscopy recalls Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in clean
Hazard: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
July 18, 2012Medical DevicesFDA
Stryker Endoscopy recalls Stryker brand Arthroscope and Hardware Set Tray (sterilization tray); Model/Part Number: 233-032-116; Product is manufac
Hazard: The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Instructions for Use"; due to the firm's evaluation that identified original validation was not done at worst case conditions.
July 4, 2012Medical DevicesFDA
Stryker Endoscopy recalls Formula Shaver Handpiece (with buttons); Formula 180 Shaver Handpiece (with button) Rx only, Made USA.: Stryker End
Hazard: Stryker Endoscopy determined that there may not have been adequate determination of required routine checks to ensure accuracy and fitness for use of specific serial numbers of Formula and 180 Shaver Hand Control.