December 9, 2015Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls Two-stage and three-stage I.V. poles are shipped as an optional accessory to a bed or stretcher. The I.V. poles are sta
Hazard: It was identified by a customer complaint that some I.V. poles were retracting/lowering from an extended position to the low height unexpectedly. An investigation was initiated which identified that the I.V. Latch Housing was nonconforming. potential hazards: " Suspended I.V. pole retracts unexpectedly " Suspended I.V. pole retracts unexpectedly with pump/bags attached
August 6, 2014Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls InTouch Critical Care Bed, Stryker Medical. Model Numbers 2130, 2131, 2140, 2141, 2151, 2152, 2155, and 2156.
Hazard: Stryker has received complaints from the field alleging instances where the In Touch beds are hoisted or lifted from the floor and the casters have fallen off. The investigations into these casters have shown that in these cases the caster has broken on the top of the caster stem where the brake rod inserts into the caster. If more than one caster on a unit is damaged, brake functions can be affec
February 12, 2014Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls The Gaymar Medi-Therm device provides a means of regulating patient temperature by supplying temperature-controlled wate
Hazard: Stryker Medical has identified that on some units of the MTA7900, there is a lack of a water temperature display while the unit is in Auto mode. While in Auto mode, only the patient temperature is displayed along with the set point. The water display will be blank during Auto mode. If the user wants to see the water temperature displayed and switches out of Auto mode to Manual mode, then back to
January 8, 2014Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls S3 Med/Surg Bed Secure II and the 3002 S3 are the beds typically found in the Med-Surg units of the hospitals. The be
Hazard: Stryker Medical has identified a potential issue associated with Secure II and S3 Med/Surg beds built between April 2nd, 2005 and July 26th, 2011. The affected beds have the potential to experience damage to the Foot End Cover and/or Power Coil Cable due to reduced clearance between the Foot End Cover and Foot End Lift Header.
January 8, 2014Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls Secure II Med/Surg Bed Secure II and the 3002 S3 are the beds typically found in the Med-Surg units of the hospitals.
Hazard: Stryker Medical has identified a potential issue associated with Secure II and S3 Med/Surg beds built between April 2nd, 2005 and July 26th, 2011. The affected beds have the potential to experience damage to the Foot End Cover and/or Power Coil Cable due to reduced clearance between the Foot End Cover and Foot End Lift Header.
November 27, 2013Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls Model FL14E Rose BedsThe Rose Bed is an AC-powered adjustable long-term care bed intended for medical purposes that co
Hazard: It was reported that the siderail may appear latched when it is not, due to lack of proper preventive maintenance. This might result in bruises and/or lacerations to a patient who is leaning against the siderails if the siderail releases. Additionally, the siderail may not effectively constrain a patient to prevent them from rolling or falling off the bed.
October 2, 2013Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls The FL23SE Bed is an AC-powered adjustable hospital bed with four built-in electric DC motors and remote controls that i
Hazard: The nurse call cable (Part Number: QDF23-0571) on the FL23SE Bed is located right under the opening of the bed head box. If an accessory (i.e. IV pole) is inserted into this slot, the circuit directing the nurse call system can be damaged.
August 21, 2013Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls IsoGel AIR 84, Support Surface. Intended to assist in improving human patient outcomes.
Hazard: Complaints were received from the field indicating that some Air Pumps were flashing and alarming when connected to the IsoGel AIR" surface upon turning the pump on and the setting it to the IsoGel AIR Mode. Upon further investigation, it was determined that the flashing/alarm being experienced was due to the air pressure being out of the expected range; indicating a blockage within the LAL tubi
January 9, 2013Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls Medical Surgical Bed, Model 3005S3EX Medical Surgical Bed The S3" MedSurg Bed, Model 3005 is intended to support and
Hazard: Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.
January 9, 2013Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls Medical Surgical Bed, Model 3002SEX Medical Surgical Bed The S3" MedSurg Bed, is intended to support and transport pa
Hazard: Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. The affected beds may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.
December 26, 2012Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls The Stryker Model 6390, Power-LOAD, is a power-loading cot fastener system designed to lift, lower, or steer ambulance c
Hazard: Model 6390 Power-LOAD systems built between September 12, 2012 and September 27, 2012 may have been built with a load switch slide that does not meet specifications causing the lifting ability of the arms to be negatively impacted.
August 29, 2012Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls Powered Mobility Offers Versatility (M¿V)TM Chair. The Powered MOV Chair models EMS, EMS24, HHC, and INO Powered MOV
Hazard: Paramed Corporation (d/b/a, Descent Control Systems), a manufacturer and distributor of patient transport devices, merged with an affiliate of Stryker Corporation in August 2009. As part of the due diligence process, Paramed's powered Mobility Offers Versatility (M¿V) chair was reviewed to identify potential compliance issues in relation to product documentation and classification.
August 29, 2012Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls Powered Mobility Offers Versatility (M¿V)TM Chair. The Powered MOV Chair models EMS, EMS24, HHC, and INO Powered MOV
Hazard: Paramed Corporation (d/b/a, Descent Control Systems), a manufacturer and distributor of patient transport devices, merged with an affiliate of Stryker Corporation in August 2009. As part of the due diligence process, Paramed's powered Mobility Offers Versatility (M¿V) chair was reviewed to identify potential compliance issues in relation to product documentation and classification.
August 29, 2012Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls Powered Mobility Offers Versatility (M¿V)TM Chair The Powered MOV Chair models EMS, EMS24, HHC, and INO Powered MOV C
Hazard: Paramed Corporation (d/b/a, Descent Control Systems), a manufacturer and distributor of patient transport devices, merged with an affiliate of Stryker Corporation in August 2009. As part of the due diligence process, Paramed's powered Mobility Offers Versatility (M¿V) chair was reviewed to identify potential compliance issues in relation to product documentation and classification.
August 29, 2012Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls Powered Mobility Offers Versatility (M¿V)TM Chair The Powered MOV Chair models EMS, EMS24, HHC, and INO Powered MOV C
Hazard: Paramed Corporation (d/b/a, Descent Control Systems), a manufacturer and distributor of patient transport devices, merged with an affiliate of Stryker Corporation in August 2009. As part of the due diligence process, Paramed's powered Mobility Offers Versatility (M¿V) chair was reviewed to identify potential compliance issues in relation to product documentation and classification.
July 4, 2012Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls Go Bed II+ Model FL 28 Ex Beds are intended for medical purposes in both general patient care as well as critical ca
Hazard: An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac
July 4, 2012Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls Stryker MA204 Bed Model (FL25E) Beds are intended for medical purposes in both general patient care as well as critic
Hazard: An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac
July 4, 2012Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls Stryker Rose Bed Model (FL14E) Beds are intended for medical purposes in both general patient care as well as critic
Hazard: An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac
July 4, 2012Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls Go Bed II+ Model FL28C Beds are intended for medical purposes in both general patient care as well as critical care
Hazard: An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac
July 4, 2012Medical DevicesFDA
Stryker Medical Division of Stryker Corporation recalls GoBed+ . Model FL20E Beds are intended for medical purposes in both general patient care as well as critical care env
Hazard: An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac