Home / Brands / Thommen Medical AG
Thommen Medical AG Recalls
2 recalls on record · May 29, 2019 to March 27, 2024 · Medical Devices
Recalls by year
19
24
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
March 27, 2024Medical DevicesFDA
Thommen Medical AG recalls impression coping, repositionable, short, screw-retained, PF 4.0
Hazard: The depth of the internal canal was not manufactured according to specifications. Thus, the impression coping cannot be firmly connected to the implant.
May 29, 2019Medical DevicesFDA
Thommen Medical AG recalls Adapter for handpiece, guided, reusable, for guide sleeve 0 4.8 mm Article number: 3.04.090 - Product Usage: The adapt
Hazard: Dental adaptor does not conform to specifications. The dental coupling of the adapter can become deformed when being inserted into the implant may make it difficult to remove the adapter from its counterpart (e.g. the handpiece).