Home / Recalls / Medical Devices
Thommen Medical AG recalls impression coping, repositionable, short, screw-retained, PF 4.0
Recalled March 27, 2024 · FDA recall Z-1322-2024 · Thommen Medical AG
Hazard
The depth of the internal canal was not manufactured according to specifications. Thus, the impression coping cannot be firmly connected to the implant. · FDA Class II
Units
20 units
Sold at
US Nationwide distribution in the states of OH and KY.
Description
The depth of the internal canal was not manufactured according to specifications. Thus, the impression coping cannot be firmly connected to the implant.
Product
impression coping, repositionable, short, screw-retained, PF 4.0
Manufacturer(s)
Thommen Medical AG
Made in
Switzerland
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →