September 10, 2025Medical DevicesFDA
Boston Scientific Corporation recalls WATCHMAN FXD Curve Access Sys Sgl, OUS, Material Number (UPN) M635TS80010; intended to provide vascular and transseptal
Hazard: Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
September 10, 2025Medical DevicesFDA
Boston Scientific Corporation recalls WATCHMAN TruSeal Access System ANT, OUS, Material Number (UPN) M635TS70040; intended to provide vascular and transseptal
Hazard: Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
September 10, 2025Medical DevicesFDA
Boston Scientific Corporation recalls WATCHMAN TruSeal Access System DBL, OUS, Material Number (UPN) M635TS70020; intended to provide vascular and transseptal
Hazard: Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
September 10, 2025Medical DevicesFDA
Boston Scientific Corporation recalls WATCHMAN TruSeal Access System SGL, OUS, Material Number (UPN) M635TS70010; intended to provide vascular and transseptal
Hazard: Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
September 10, 2025Medical DevicesFDA
Baxter Healthcare Corporation recalls Novum IQ Syringe Pump, Product Code 40700BAX, infusion pump
Hazard: Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).
September 10, 2025Medical DevicesFDA
Baxter Healthcare Corporation recalls Novum IQ Syringe Pump, Product Code 40800BAXUS, infusion pump
Hazard: Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).
September 10, 2025Medical DevicesFDA
Baxter Healthcare Corporation recalls Novum IQ LVP (Large Volume Pump), Product Code 40700BAXUS, infusion pump
Hazard: Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).
September 3, 2025Medical DevicesFDA
MEDLINE INDUSTRIES, LP - Northfield recalls Medline Kits containing BD SureStep Foley Tray Systems: SKU/Description: CDS984076J GYN ROBOTIC DYNJG001004A LAVH
Hazard: Kits contain SureStep Foley Tray Systems recalled by BD that include incorrect product information insert sheets. If a mislabeled tray is used, patient risk ranges from minor injury to injury that requires medical intervention, such as infection or local/systemic allergic reaction (anaphylaxis).
September 3, 2025Medical DevicesFDA
CareFusion 303, Inc. recalls BD Pyxis Product Name / UDI-DI code / Catalog No. BD Pyxis" MedStation" 4000 Main / 10885403512629 / 303 BD Pyxis"
Hazard: Antivirus software was not consistently installed on impacted devices during the implementation process.
September 3, 2025Medical DevicesFDA
Diagnostica Stago, Inc. recalls STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers o
Hazard: Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
September 3, 2025Medical DevicesFDA
Diagnostica Stago, Inc. recalls STA-Liquid Anti-Xa 8. Model/Catalog Number: 00322US. The STA - Liquid Anti-Xa kits are intended for use with STA-R,
Hazard: Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
September 3, 2025Medical DevicesFDA
Diagnostica Stago, Inc. recalls STA-Liquid Anti-Xa 4. Model/Catalog Number: 00311US. The STA - Liquid Anti-Xa kits are intended for use with STA-R, S
Hazard: Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
September 3, 2025Medical DevicesFDA
Diagnostica Stago, Inc. recalls STA Satellite. Product Code: All references. All software versions. The STA Satellite is a fully automated benchtop a
Hazard: Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
September 3, 2025Medical DevicesFDA
American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ANTERIOR HIP PACK, MODEL/ITEM NUMBER AMAH4
Hazard: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
September 3, 2025Medical DevicesFDA
American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ROBOT HYST PACK, MODEL/ITEM NUMBER GRRB10E
Hazard: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
September 3, 2025Medical DevicesFDA
American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAPAROSCOPY PACK, MODEL/ITEM NUMBER EVLP61
Hazard: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
September 3, 2025Medical DevicesFDA
American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) BORN ON ARRIVAL KIT, Model/Item Number LLB
Hazard: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
September 3, 2025Medical DevicesFDA
American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) SFCH CRANI PACK (PS 132432), MODEL/ITEM NU
Hazard: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
September 3, 2025Medical DevicesFDA
American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) A-C SUPPLEMENT PACK - 205946, Model/Item
Hazard: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
September 3, 2025Medical DevicesFDA
American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) CYSTO, Model/Item Number ASCY24A; 2) CYS
Hazard: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
September 3, 2025Medical DevicesFDA
American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAP CHOLE PACK, Model/Item Number CILC69L;
Hazard: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
September 3, 2025Medical DevicesFDA
American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) HEART PACK - 205947, MODEL/ITEM NUMBER ANC
Hazard: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
September 3, 2025Medical DevicesFDA
Zhejiang Luyao Electronics Technology Co., Ltd. recalls Brand Name: HOMEDICS Product Name: Air Compression Leg Massager Model/Catalog Number: SR-160HJ Product Description: P
Hazard: Regulatory documentation error and misleading color packaging claim.
September 3, 2025Medical DevicesFDA
C.R. Bard Inc recalls Bardex Lubri-Sil I.C. 400-series Temperature Sensing Silver/Hydrogel Coated All-Silicone Foley Catheter 5cc Balloon, REF
Hazard: Foley catheter may have an obstructed lumen, which may result in an inability to drain the bladder, of urine and may include pelvic or abdominal pain, flank pain, bladder dysfunction, obstructive uropathy, acute kidney injury and/or hydronephrosis.
September 3, 2025Medical DevicesFDA
Medtronic Xomed, Inc. recalls IPC Powerease System REF 2300000 UDI-DI: 00613994448705 00643169406834. The IPC POWEREASE System is indicated for
Hazard: Due to out of the box wobble of the driver.
September 3, 2025Medical DevicesFDA
MEDLINE INDUSTRIES, LP - Northfield recalls Medline ReNewal Reprocessed St. Jude Medical Response and Supreme Diagnostic Electrophysiology Catheter Models 401150RH
Hazard: Specific lots of reprocessed electrophysiology catheters may contain small residual particulates.
September 3, 2025Medical DevicesFDA
BD SWITZERLAND SARL recalls BD PhaSeal Injector Luer (N30C), REF: 515001 BD PhaSeal Injector Luer Lock (N35), REF: 515003 BD PhaSeal Injector Luer
Hazard: Closed system drug transfer devices were shipped to the U.S. market without a U.S. version of the instructions for use, which could be an issue if a healthcare professional is prompted to review it.
September 3, 2025Medical DevicesFDA
MEDLINE INDUSTRIES, LP - Northfield recalls The Medline General Surgery Tray is customized to meet requirements of the hospitals and health care professionals utili
Hazard: Affected kits contain recalled Stryker Color Cuff Sterile Disposable Tourniquet Cuffs. Stryker Instruments initiated this recall due to a potential manufacturing defect in which the flange may detach from the bladder at the welded connection point. This detachment compromises the cuff's ability to maintain adequate air pressure, potentially resulting in tourniquet failure and increased risk of patient hemorrhaging.
September 3, 2025Medical DevicesFDA
Bausch & Lomb Surgical, Inc. recalls Akreos, SKU: AO60P0300. Akreos intraocular lenses
Hazard: An off-power error was identified, which resulted in a diopter of 16, rather than the intended 3 diopter. Implantation of affected intraocular lens could result in functional visual impairment or reduction in visual acuity which could result in the need for optical or surgical correction.
September 3, 2025Medical DevicesFDA
Orascoptic Surgical Acuity recalls Orascoptic Superior Visualization Phantom Nose Pad Accessory Kit ( nose pad and screw) 1 pk. and 20 pk.
Hazard: Due to complaints, loupe nose pads containing nickel may cause an allergic reaction.