RecallWatchUSA

Home / Recalls / Medical Devices

Abbott Point Of Care Inc. recalls Abbott i-STAT EG7+ cartridge-IVD. Intended for the quantitative measurement of sodium, potassium, ionized calcium, pH,

Recalled January 27, 2021 · FDA recall Z-0881-2021 · Abbott Point Of Care

Hazard
Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y. · FDA Class II
Units
1,521,800 cartridges US and 2,641,053 cartridges OUS
Sold at
Nationwide

Description

Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.

Product
Abbott i-STAT EG7+ cartridge-IVD. Intended for the quantitative measurement of sodium, potassium, ionized calcium, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood. List Number: 03P76-25
Manufacturer(s)
Abbott Point Of Care Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Abbott Point Of Care Inc. recalls Abbott i-STAT EG7+ cartridge-IVD. Intended for the quantitative measurement of sodium, potassium, ionized calcium, pH, | RecallWatchUSA