Home / Brands / Abbott Point Of Care
Abbott Point Of Care Recalls
31 recalls on record · July 18, 2012 to May 6, 2026 · Medical Devices
Recalls by year
12
13
14
15
18
20
21
25
26
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
May 6, 2026Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT G3+ cartridge; List Number: 03P78-26;
Hazard: Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
May 6, 2026Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT EG6+ cartridge; List Number: 03P77-25;
Hazard: Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
May 6, 2026Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT EG7+ cartridge; List Number: 03P76-25;
Hazard: Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
September 24, 2025Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT EG6+ cartridge. List Number: 03P77-25.
Hazard: Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.
September 24, 2025Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT EG7+ cartridge. List Number: 03P76-25.
Hazard: Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.
September 24, 2025Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT CG8+ cartridge. List Number: 03P88-25. 510(k) numbers: K894914, K940918, K223710, K230275, K230285, K230300
Hazard: Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.
September 24, 2025Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT CG4+ cartridge (white). List Number: 03P85-25.
Hazard: Lack of a 510(k) premarket clearance for the i-STAT CG4+ cartridge to reflect updated sample type for lactate and measurement ranges for pH and PO2.
January 27, 2021Medical DevicesFDA
Abbott Point Of Care Inc. recalls Abbott i-STAT EG7+ cartridge-IVD. Intended for the quantitative measurement of sodium, potassium, ionized calcium, pH,
Hazard: Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.
January 27, 2021Medical DevicesFDA
Abbott Point Of Care Inc. recalls Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose
Hazard: Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.
February 26, 2020Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT G3+ cartridge (blue), List No. 03P78-50, UDI (01) 00054749002283 - Product Usage: The i-STAT G3+ cartridge may be
Hazard: Marketed without a 510(k). Abbott Point of Care has determined that it does not have sufficient clinical evidence to fully characterize the performance of the blue i-STAT G3+ cartridges and therefore will no longer distribute the i-STAT G3+ (blue) cartridge in the United States as of May 1, 2020.
February 26, 2020Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT CG4+ cartridge (blue), List No. 03P85-50. UDI (01) 00054749002269 - Product Usage: The i-STAT CG4+ cartridge is i
Hazard: The I-STAT BLUE CHEM8+ and CG4+ cartridges are not FDA cleared and do not have CLIA waived status. CLIA recognizes two types of laboratory tests: waived and non-waived. Different CLIA standards apply to waived and non-waived tests as well as the facilities that perform these types of tests. The i-STAT CHEM8+ cartridge contains nine measured assays (sodium, potassium, chloride, blood urea nitrogen (BUN), ionized calcium, TCO2, glucose, creatinine and hematocrit) and is categorized as a CLIA waived test for venous whole blood samples, meaning this cartridge could be used in a facility with a CLIA Certificate of Waiver as a waived test.
February 26, 2020Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT CHEM8+ cartridges (blue), List No. 09P31-25. UDI (01) 00054749001910 - Product Usage: The i-STAT CHEM8+ cartridge
Hazard: The I-STAT BLUE CHEM8+ and CG4+ cartridges are not FDA cleared and do not have CLIA waived status. CLIA recognizes two types of laboratory tests: waived and non-waived. Different CLIA standards apply to waived and non-waived tests as well as the facilities that perform these types of tests. The i-STAT CHEM8+ cartridge contains nine measured assays (sodium, potassium, chloride, blood urea nitrogen (BUN), ionized calcium, TCO2, glucose, creatinine and hematocrit) and is categorized as a CLIA waived test for venous whole blood samples, meaning this cartridge could be used in a facility with a CLIA Certificate of Waiver as a waived test.
January 8, 2020Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT cTnI cartridge
Hazard: i-STAT¿ cTnI blue cartridges may generate a higher than expected number of quality check codes. The observed rate of quality check codes could be in the 3 to 5% range.
December 26, 2018Medical DevicesFDA
Abbott Point Of Care Inc. recalls Abbott i-STAT CG8+ Cartridges intended for use in the in vitro quantification of potassium in arterial, venous, or capil
Hazard: Inaccurate potassium results may be generated. When testing with control fluid and/or patient samples with low or reference range potassium concentration, elevated potassium results may be generated. When testing with samples with a high concentration of potassium, decreased potassium results may be generated
March 14, 2018Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054
Hazard: Issues resulting from upgrade to software version 2.8: (1) Location, operator, stored patient lists will not update, and (2) Customized Reference Ranges, Action Ranges, and Custom Reportable Ranges are reset to factory default values. No erroneous results are generated as a result of this issue.
August 19, 2015Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-Stat PT/INR Cartridges List number 03P89-24 The cartridges are packaged in sealed pouches, one per pouch. There ar
Hazard: Certain lots of PT/INR cartridges may generate a higher than typical number of PT/INR star outs potentially delaying the generation of patient results for PT/INR.
April 2, 2014Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT G3+ Cartridge pH, pCO2 and pO2 measurements are used in the diagnosis, monitoring and treatment of respiratory
Hazard: Abbott Point of Care has determined that some individual patient results generated with the i-STAT G3+ cartridge lot N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results.
December 11, 2013Medical DevicesFDA
Abbott Point Of Care Inc. recalls Abbott i-STAT PT/INR cartridges Abbott Point of Care Inc., Princeton, NJ 08540 The i-Stat PT a prothrombin time test
Hazard: Abbott Point of Care Inc (APOC) has determined that i-Stat PT/INR cartridges have the potential to exhibit incorrectly elevated INR.
November 14, 2012Medical DevicesFDA
Abbott Point Of Care Inc. recalls Cliniqa Liquid QCTM Cardiac Marker Control for i-STAT Level 1 (List Number 06F12-09; Lot Number 071039) Cliniqa Liqui
Hazard: Abbott Point of Care (APOC) has determined that for the Cliniqa Liquid QC Cardiac Marker Control for i-STAT Level 1 (List #06F12-09; Lot Number 071039); i-STAT Level 2 (List #06F12-10; Lot Number 081039); and i-STAT Level 3 (List #06F12-11; Lot Number 091039) results may be generated below the lower value assignment range for some vials.
August 29, 2012Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT cTnI cartridges Product Usage: The i-STAT cTnl test is an in vitro diagnostic test for the quantitative mea
Hazard: Abbott Point of Care has determined that i-Stat CTNL cartridges have the potential to exhibit incorrectly elevated cardiac troponin I results. This recall was expanded on July 19 2012 to include additional information regarding user documentation.
August 15, 2012Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-Stat CK-MP cartridges Abbott Point of Care Inc. The i-Stat CKMB test is an in vitro diagnostic test for the quanti
Hazard: The frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure.
August 15, 2012Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-Stat BNP cartridges Abbott Point of Care Inc. The i-Stat BNP test is an in vitro diagnostic test for the quantita
Hazard: The frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure.
August 15, 2012Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-Stat cTnl cartridges Abbott Point of Care Inc. The i-Stat cTnl test is an in vitro diagnostic test for the quant
Hazard: The frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure.
August 15, 2012Medical DevicesFDA
Abbott Point Of Care Inc. recalls I-STAT PT/INR cartridge Abbott Point of Care Inc. Abbott Park, IL 60064 USA This cartridge is useful for monitoring
Hazard: The i-STAT PT/INR cartridges have the potential to exhibit incorrectly elevated results.
August 8, 2012Medical DevicesFDA
Abbott Point Of Care Inc. recalls I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quanti
Hazard: i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer.
July 18, 2012Medical DevicesFDA
Abbott Point Of Care Inc. recalls Military Kit Portable Clinical Analyzer ; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests f
Hazard: Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.
July 18, 2012Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT1 Abaxis Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassi
Hazard: Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.
July 18, 2012Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-STAT1 Abaxis Analyzer; Abbott Point of Care, Princeton, NJ 08540
Hazard: Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.
July 18, 2012Medical DevicesFDA
Abbott Point Of Care Inc. recalls Fuso Analyzer 300F; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassium, i
Hazard: Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.
July 18, 2012Medical DevicesFDA
Abbott Point Of Care Inc. recalls i-Stat Portable Clinical Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodiu
Hazard: Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.