Abbott Point Of Care Inc. recalls I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quanti

Recalled August 8, 2012 · FDA recall Z-2133-2012 · Abbott Point Of Care

Hazard
i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer. · FDA Class II
Units
52,750 cartridges
Sold at
USA ( nationwide )

Description

i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer.

Product
I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium, potassium, ionized calcium, oxygen, carbon dioxide, partial pressure in arterial, venous, or capillary whole blood.
Manufacturer(s)
Abbott Point Of Care Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov