RecallWatchUSA

Home / Recalls / Medical Devices

Abbott Point Of Care Inc. recalls i-STAT EG6+ cartridge; List Number: 03P77-25;

Recalled May 6, 2026 · FDA recall Z-2000-2026 · Abbott Point Of Care

Hazard
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems. · FDA Class II
Units
30775 units (25625 US, 5150 OUS)
Sold at
Worldwide distribution - US Nationwide and the countries of Canada, AUSTRIA, BANGLADESH, BELGIUM, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, LUXEMBOURG, MAURITIUS, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM.

Description

Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.

Product
i-STAT EG6+ cartridge; List Number: 03P77-25;
Manufacturer(s)
Abbott Point Of Care Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Abbott Point Of Care Inc. recalls i-STAT EG6+ cartridge; List Number: 03P77-25; | RecallWatchUSA