RecallWatchUSA

Home / Recalls / Medical Devices

Advanced Sterilization Products recalls EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reproces

Recalled June 17, 2015 · FDA recall Z-1714-2015 · Advanced Sterilization Products

Hazard
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope. · FDA Class II
Units
825 units
Sold at
Worldwide Distribution-US (nationwide) and the countries of Australia, Singapore, Canada, and Thailand.

Description

Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.

Product
EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 ¿C) semi-critical endoscopes.
Manufacturer(s)
Advanced Sterilization Products

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Advanced Sterilization Products recalls EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reproces | RecallWatchUSA