Advanced Sterilization Products Recalls

36 recalls on record · August 8, 2012 to February 22, 2017 · Medical Devices

Recalls by year

12
13
14
15
17

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

February 22, 2017Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD NX Sterilization System, Product Code: 10033, 10033-002
Hazard: Advanced Sterilization Products (ASP) has identified that in a rare sequence of events following an external power outage, the system door interlock can be become disengaged during system reboot for a 20-second duration.
January 18, 2017Medical DevicesFDA
Advanced Sterilization Products recalls EVOTECH Endoscope Cleaner and Reprocessor System 208V, Product Code: 50004 The EVOTECH ECR Endoscope Cleaner and Repr
Hazard: Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor System because the hose may come loose and cause a leak.
September 23, 2015Medical DevicesFDA
Advanced Sterilization Products recalls CIDEX OPA Solution, P/N 20390 for use as a high level disinfectant for reprocessing heat sensitive medical devices.
Hazard: Advanced Sterilization Products (ASP) is reminding all CIDEX OPA Solution customers to follow the warnings and rinsing procedure in their current labelling around the criticality of proper rinsing of medical device following disinfection with CIDEX OPA according to the CIDEX OPA Solution IFU (Instruction For Use).
September 2, 2015Medical DevicesFDA
Advanced Sterilization Products recalls CIDEX Activated Dialdehyde Solution, 4 bottles/case, P/N: 2266
Hazard: Advanced Sterilization Products (ASP) is recalling the CIDEX Activated Dialdehyde Solution because it may lack the adequate data to support the entire 24-month shelf life.
August 19, 2015Medical DevicesFDA
Advanced Sterilization Products recalls Wall Chart for STERRAD 100NX System, P/N 10104. The STERRAD 100NX Sterilizer is designed for sterilization of both me
Hazard: Advanced Sterilization PRoducts (ASP) is recalling the Wall Chart for STERRAD 100NX System because it incorrectly states "APTIMAX Instrument Trays can be enclosed in either polypropylene wrap or in Tyvek Pouches with STERRAD Chemical Indicators."
June 17, 2015Medical DevicesFDA
Advanced Sterilization Products recalls EVOTECH Endoscope Cleaner and Reprocessing System 220-230V, P/N 50004-002. The ASP EVOTECH ECR Endoscope Cleaner and
Hazard: Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.
June 17, 2015Medical DevicesFDA
Advanced Sterilization Products recalls EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reproces
Hazard: Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.
April 1, 2015Medical DevicesFDA
Advanced Sterilization Products recalls Tyvek Pouch/Roll with STERRAD Chemical Indicator, P/N 12425
Hazard: Advanced Sterilization Products (ASP) is recalling the Tyvek Rolls with STERRAD Chemical Indicator because they were shipped to customers after its expiration date of May 18, 2013.
December 31, 2014Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD 100NX, Product Code 10144 The STERRAD Cassettes are used on STERRAD Sterilization System which is a low tempe
Hazard: Advanced Sterilization Products is recalling the STERRAD System cassettes because the the leak indicator on the STERRAD System cassettes packaging may not function properly after exposure to hydrogen peroxide.
December 31, 2014Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD NX, Product Code 10133 The STERRAD Cassettes are used on STERRAD Sterilization System which is a low temperat
Hazard: Advanced Sterilization Products is recalling the STERRAD System cassettes because the the leak indicator on the STERRAD System cassettes packaging may not function properly after exposure to hydrogen peroxide.
December 31, 2014Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD 200, Product Code 10118 The STERRAD Cassettes are used on STERRAD Sterilization System which is a low tempera
Hazard: Advanced Sterilization Products is recalling the STERRAD System cassettes because the the leak indicator on the STERRAD System cassettes packaging may not function properly after exposure to hydrogen peroxide.
December 31, 2014Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD 100S, Product Code 10113. The STERRAD Cassettes are used on STERRAD Sterilization System which is a low tempe
Hazard: Advanced Sterilization Products is recalling the STERRAD System cassettes because the the leak indicator on the STERRAD System cassettes packaging may not function properly after exposure to hydrogen peroxide.
December 31, 2014Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD 50, Product Code 10114. The STERRAD Cassettes are used on STERRAD Sterilization System which is a low tempera
Hazard: Advanced Sterilization Products is recalling the STERRAD System cassettes because the the leak indicator on the STERRAD System cassettes packaging may not function properly after exposure to hydrogen peroxide.
January 22, 2014Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD 100S Sterilization System, Product Code: 10101 The STERRAD 100S Sterilization System is a general purpose, low
Hazard: Advanced Sterilization Products (ASP) is recalling the STERRAD NX, STERRAD 50, and STERRAD 100S Sterilization Systems because some healthcare workers detected odors or smells possibly emanating from STERRAD Systems.
January 22, 2014Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD 50 Sterilization System, Product Code: 10050 The STERRAD 50 Sterilization System is a general purpose, low temp
Hazard: Advanced Sterilization Products (ASP) is recalling the STERRAD NX, STERRAD 50, and STERRAD 100S Sterilization Systems because some healthcare workers detected odors or smells possibly emanating from STERRAD Systems.
January 22, 2014Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD NX Sterilization System, Product Code: 10033 The STERRAD NX Sterilization System is a general purpose, low te
Hazard: Advanced Sterilization Products (ASP) is recalling the STERRAD NX, STERRAD 50, and STERRAD 100S Sterilization Systems because some healthcare workers detected odors or smells possibly emanating from STERRAD Systems.
January 8, 2014Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD Cyclesure Biological Indicator (BI, Part No. 14324. Intended to be used as a standard method for frequent mon
Hazard: Advanced Sterilization Products (ASP) is recalling the STERRAD Cyclesure Biological Indicator (BI) because ASP has identified a microscopic intermittent and low frequency flaw in the outer vial of the STERRAD Cyclesure 24 Biological Indicator.
December 11, 2013Medical DevicesFDA
Advanced Sterilization Products recalls STERITITE Containers: STERITITE Container Kit, 1/4 size, 3" high, product code: 13902; STERITITE Mini Mid Size, 3" high,
Hazard: Advanced Sterilization Products (ASP) is recalling the SteriTite containers because they do not have consistent data to show that sterility assurance level of 10(-6) is achieved during the half-cycle test runs on new, unprocessed containers.
December 4, 2013Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD NX, Product Code: 10033 The STERRAD Steri lization System is a low temperature general purpose sterilizer use
Hazard: Advanced Sterilization Products (ASP) is recalling the STERRAD NX and STERRAD 100NX sterilization systems because ASP has determined that the bronchoscopes may contain a polyurethane material in the suction channel location that is not listed in the current claims for interior channels (lumens) for processing in STERRAD systems.
December 4, 2013Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD 100NX, Product Code: 10104 The STERRAD Steri lization System is a low temperature general purpose sterilizer
Hazard: Advanced Sterilization Products (ASP) is recalling the STERRAD NX and STERRAD 100NX sterilization systems because ASP has determined that the bronchoscopes may contain a polyurethane material in the suction channel location that is not listed in the current claims for interior channels (lumens) for processing in STERRAD systems.
November 20, 2013Medical DevicesFDA
Advanced Sterilization Products recalls Sterrad 100NX, Product Code: 10104. Used to sterilize heat and moisture sensitive reusable medical devices.
Hazard: Advanced Sterilization Products (ASP) is recalling the Sterrad 100NX Sterilizer some healthcare workers detected odors/smells emanating from the 100NX Sterilizers, and it may be related to normal breakdown of the oil used in the Sterrad 200 vacuum pumps.
October 30, 2013Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to ster
Hazard: Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.
October 30, 2013Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD 200 Sterilizer (Single Door), Part Number: 10201. STERRAD 200 Sterilizer (Double Door), Part Number: 10202.
Hazard: Advanced Sterilization Products (ASP) is recalling the STERRAD 200 Sterilization System because it may emit an odor or smell into the surrounding environment that is not typical to normal operating conditions.
September 18, 2013Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD 100NX Cassettes, Product Code: 10144 The STERRAD Sterilization System is a low temperature general purpose st
Hazard: The recall of the STERRAD 100NX Cassettes was initiated because Advanced Sterilization Products (ASP) does not have adequate data to support the effectiveness of the packaging's leak indicator for the entire duration of the STERRAD 100 NX System cassette's 15 month shelf life.
May 8, 2013Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD CYCLESURE 24 Biological Indicator (BI) The STERRAD CYCLESURE 24 Biological Indicator (BI), is intended to be
Hazard: Advanced Sterilization Products is recalling certain lots of STERRAD CYCLESURE 24 Biological Indicator product because it may not have adequate data to support the entire duration of the labeled shelf-life.
April 24, 2013Medical DevicesFDA
Advanced Sterilization Products recalls EVOTECH Endoscope Cleaner and Reprocessor System, P/N 50004. Designed to automatically clean and high-level disinfect
Hazard: Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because they have received reports of smoke emanating from the system due to a circuit board malfunction that may occur in the Channel Control Manifold sub-assembly.
April 24, 2013Medical DevicesFDA
Advanced Sterilization Products recalls EVOTECH Endoscope Cleaner & Reprocessor System, P/N 50004 Designed to automatically clean and high-level disinfect fl
Hazard: Advanced Sterilization Products (ASP) has determined that a limited number of EVOTECH ECRs manufactured between 2011 and 2012 did not receive sufficient High Potential (HIPOT) DC voltage testing.
April 10, 2013Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD 200 Sterilization System, P/N 10201 Product Usage: The STERRAD 200 Sterilization System is a general purpose
Hazard: The recall was initiated because Advanced Sterilization Products (ASP) has determined that the STERRAD 100NX and STERRAD 200 Sterilization Systems may emit an odor or smell into the surrounding environment that is not typical to normal operating conditions.
April 10, 2013Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD 100NX Sterilization System, P/N 10104 Product Usage: The STERRAD 100NX Sterilization System is a general pur
Hazard: The recall was initiated because Advanced Sterilization Products (ASP) has determined that the STERRAD 100NX and STERRAD 200 Sterilization Systems may emit an odor or smell into the surrounding environment that is not typical to normal operating conditions.
April 3, 2013Medical DevicesFDA
Advanced Sterilization Products recalls STERRAD 200 System Carriage, P/N 10208. Used as a tool to expedite loading and unloading processed items from a STERR
Hazard: Advanced Sterilization Products (ASP) is recalling the STERRAD 200 System Carriage because it can potentially become dislodged when an excessive load is applied.