Home / Recalls / Medical Devices
Advanced Sterilization Products recalls EVOTECH Endoscope Cleaner and Reprocessor System, P/N 50004. Designed to automatically clean and high-level disinfect
Recalled April 24, 2013 · FDA recall Z-1150-2013 · Advanced Sterilization Products
Hazard
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because they have received reports of smoke emanating from the system due to a circuit board malfunction that may occur in the Channel Control Manifold sub-assembly. · FDA Class II
Units
761 units
Sold at
Nationwide Distribution
Description
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because they have received reports of smoke emanating from the system due to a circuit board malfunction that may occur in the Channel Control Manifold sub-assembly.
Product
EVOTECH Endoscope Cleaner and Reprocessor System, P/N 50004. Designed to automatically clean and high-level disinfect flexible, submersible video or fiber-optic endoscopes.
Manufacturer(s)
Advanced Sterilization Products
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →