Advanced Sterilization Products recalls EVOTECH Endoscope Cleaner & Reprocessor System, P/N 50004 Designed to automatically clean and high-level disinfect fl

Recalled April 24, 2013 · FDA recall Z-1123-2013 · Advanced Sterilization Products

Hazard
Advanced Sterilization Products (ASP) has determined that a limited number of EVOTECH ECRs manufactured between 2011 and 2012 did not receive sufficient High Potential (HIPOT) DC voltage testing. · FDA Class II
Units
273 units
Sold at
Nationwide Distribution-USA including DC and the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI.

Description

Advanced Sterilization Products (ASP) has determined that a limited number of EVOTECH ECRs manufactured between 2011 and 2012 did not receive sufficient High Potential (HIPOT) DC voltage testing.

Product
EVOTECH Endoscope Cleaner & Reprocessor System, P/N 50004 Designed to automatically clean and high-level disinfect flexible, submersible video or fiber-optic endoscopes.
Manufacturer(s)
Advanced Sterilization Products