Home / Recalls / Medical Devices
Advanced Sterilization Products recalls EVOTECH Endoscope Cleaner and Reprocessing System 220-230V, P/N 50004-002. The ASP EVOTECH ECR Endoscope Cleaner and
Recalled June 17, 2015 · FDA recall Z-1715-2015 · Advanced Sterilization Products
Hazard
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope. · FDA Class II
Units
44 units
Sold at
Worldwide Distribution-US (nationwide) and the countries of Australia, Singapore, Canada, and Thailand.
Description
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.
Product
EVOTECH Endoscope Cleaner and Reprocessing System 220-230V, P/N 50004-002. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 ¿C) semi-critical endoscopes.
Manufacturer(s)
Advanced Sterilization Products
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →