Home / Recalls / Medical Devices
Augustine Temperature Management, LLC recalls Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hyp
Recalled February 26, 2025 · FDA recall Z-1198-2025 · Augustine Temperature Management
Hazard
There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU. · FDA Class II
Units
1 unit
Sold at
US Nationwide distribution in the states of TX, AL, NV; AU, NZ.
Description
There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
Product
Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hypothermia
Manufacturer(s)
Augustine Temperature Management, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →