Home / Brands / Augustine Temperature Management
Augustine Temperature Management Recalls
13 recalls on record · December 25, 2024 to February 26, 2025 · Medical Devices
Recalls by year
24
25
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
February 26, 2025Medical DevicesFDA
Augustine Temperature Management, LLC recalls Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 89 cm (35"), REF US530, to prevent or treat hypo
Hazard: There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
February 26, 2025Medical DevicesFDA
Augustine Temperature Management, LLC recalls Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hyp
Hazard: There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
February 26, 2025Medical DevicesFDA
Augustine Temperature Management, LLC recalls Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypo
Hazard: There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
February 26, 2025Medical DevicesFDA
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming
Hazard: Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .
February 26, 2025Medical DevicesFDA
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A136 used in conjunction with Warming
Hazard: Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .
January 22, 2025Medical DevicesFDA
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) WC77V;
Hazard: HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
January 22, 2025Medical DevicesFDA
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) WC71V;
Hazard: HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
January 22, 2025Medical DevicesFDA
Augustine Temperature Management, LLC recalls Hillrom Temperature Management Controller, MP, Model # 2083516; used in conjunction with Warming Blankets and Mattresses
Hazard: HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
January 22, 2025Medical DevicesFDA
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) Model # W
Hazard: HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
January 22, 2025Medical DevicesFDA
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) Model # W
Hazard: HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
December 25, 2024Medical DevicesFDA
Augustine Temperature Management, LLC recalls Hillrom Temperature Management Controller, MP, Model # 2083516; used in conjunction with Warming Blankets and Mattresses
Hazard: Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST".
December 25, 2024Medical DevicesFDA
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Multiport, Models: 1) Model # WC7
Hazard: Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST".
December 25, 2024Medical DevicesFDA
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) Model # W
Hazard: Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST".