Augustine Temperature Management, LLC recalls Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 89 cm (35"), REF US530, to prevent or treat hypo

Recalled February 26, 2025 · FDA recall Z-1199-2025 · Augustine Temperature Management

Hazard
There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU. · FDA Class II
Units
3 units
Sold at
US Nationwide distribution in the states of TX, AL, NV; AU, NZ.

Description

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

Product
Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 89 cm (35"), REF US530, to prevent or treat hypothermia
Manufacturer(s)
Augustine Temperature Management, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov