Beckman Coulter Inc. recalls Access 2 Immunoassay System, Remanufactured, Catalog Number 386220. In vitro diagnostic device used for the quantitat

Recalled January 10, 2018 · FDA recall Z-0309-2018 · Beckman Coulter

Hazard
Potential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment. · FDA Class III
Units
3 units
Sold at
Worldwide Distribution -- US, Argentina, Canada, Chile, China, Columbia, Costa Rica, France, Georgia, Germany, Ghana, Greece, Hong Kong, India, Italy, Japan, Kazakhstan, Mexico, Morocco, Myanmar, Nepal, Philippines, Poland, Russia, Rwanda, Saudi Arabia, Slovakia, South Africa, South Korea, Thailand, Turkey, United Kingdom, and Vietnam.

Description

Potential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment.

Product
Access 2 Immunoassay System, Remanufactured, Catalog Number 386220. In vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.
Manufacturer(s)
Beckman Coulter Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov