RecallWatchUSA

Home / Recalls / Medical Devices

Beckman Coulter, Inc. recalls Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney,

Recalled March 12, 2025 · FDA recall Z-1249-2025 · Beckman Coulter

Hazard
Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased results, which may lead to improper diagnosis or repeat testing. · FDA Class II
Units
498 US; 1576 OUS
Sold at
Domestic distribution to AL AR AZ CA CT FL ID IL IN KS MA MI NC NE NJ NY OH OR PA TN TX UT WA WV. International distribution to Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Korea, Republic of, Netherlands, Philippines, Russian Federation, Saudi Arabia, Slovakia, Spain, United Kingdom of Great Britain and Northern Ireland

Description

Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased results, which may lead to improper diagnosis or repeat testing.

Product
Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney, is the principal factor regulating red blood cell production (erythropoiesis) in mammals. Renal production of EPO is regulated by changes in oxygen availability. Under conditions of hypoxia, the level of EPO in the circulation increases and this leads to increased production of red blood cells.
Manufacturer(s)
Beckman Coulter, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Beckman Coulter, Inc. recalls Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney, | RecallWatchUSA