Home / Recalls / Medical Devices
Beckman Coulter Inc. recalls Access Immunoassay Systems Reaction Vessels. The Access Immunoassay System is an in vitro diagnostic device used for
Recalled December 25, 2013 · FDA recall Z-0519-2014 · Beckman Coulter
Hazard
Beckman Coulter is initiating a recall for the Access Immunoassay Systems Reaction Vessels, including the Acess/Access 2 Immunoassay Systems, SYNCRON LXi 725 Clinical System, and UniCel DxC 600i SYNCHRON Access Clinical System because Beckman Coulter has observed an intermittent increase in Access System "Wash Carousel Motion Errors" with specific lots of Access System Reaction Vessels (RVs). I · FDA Class II
Units
16,154 units
Sold at
Worldwide Distribution - USA (nationwide) and Internationally to Canada, Japan, India, Hong Kong, Guatemala, China, Brazil, Brunei Darussalam, Costa Rica, El Salvador, JordanMalaysia, Mexico, Panama, Philippines, South Africa, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates, Venezuela, and Viet Nam.
Description
Beckman Coulter is initiating a recall for the Access Immunoassay Systems Reaction Vessels, including the Acess/Access 2 Immunoassay Systems, SYNCRON LXi 725 Clinical System, and UniCel DxC 600i SYNCHRON Access Clinical System because Beckman Coulter has observed an intermittent increase in Access System "Wash Carousel Motion Errors" with specific lots of Access System Reaction Vessels (RVs). I
Product
Access Immunoassay Systems Reaction Vessels. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
Manufacturer(s)
Beckman Coulter Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →