RecallWatchUSA

Home / Recalls / Medical Devices

Beckman Coulter, Inc. recalls Access Testosterone assay, Catalog Number 33560

Recalled June 11, 2025 · FDA recall Z-1901-2025 · Beckman Coulter

Hazard
Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results. · FDA Class II
Units
70 units
Sold at
US State: FL. UAE, Vietnam

Description

Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.

Product
Access Testosterone assay, Catalog Number 33560
Manufacturer(s)
Beckman Coulter, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Beckman Coulter, Inc. recalls Access Testosterone assay, Catalog Number 33560 | RecallWatchUSA