Beckman Coulter, Inc. recalls Access Total T4, REF 33800, IVD, thyroxine reagent assay.

Recalled March 27, 2024 · FDA recall Z-1303-2024 · Beckman Coulter

Hazard
Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results. · FDA Class II
Units
64,009 kits
Sold at
Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Bahrain, Bangladesh, Bolivia, Brazil, Canada, China, Croatia, Czech Republic, Denmark, Egypt, El Salvador, Georgia, Greece, Hong Kong, India, Kazakhstan, Mexico, Namibia, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Taiwan, Thailand, Turkey, U.A.E, Vietnam, and Zimbabwe.

Description

Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.

Product
Access Total T4, REF 33800, IVD, thyroxine reagent assay.
Manufacturer(s)
Beckman Coulter, Inc.