Home / Recalls / Medical Devices
Beckman Coulter Inc. recalls AutoMate 1250, Sorter with Aliquot Module - UDI: 15099590364106; Model #ODL25125 Automate 2550, High Speed Sorter with
Recalled November 27, 2019 · FDA recall Z-0492-2020 · Beckman Coulter
Hazard
Beckman Coulter has identified a case where a customer using the microltiterplate kit option without a sorting drive encountered a cross-contamination of samples when a connection time-out occurred between the system and Laboratory Information System (LIS) . · FDA Class II
Units
83 automate systems
Sold at
US: NY, VA OUS: Austria Belgium Czech Republic France Germany Israel Italy Lithuania Mayotte Romania Slovakia South Africa Spain Sweden Switzerland United Kingdom
Description
Beckman Coulter has identified a case where a customer using the microltiterplate kit option without a sorting drive encountered a cross-contamination of samples when a connection time-out occurred between the system and Laboratory Information System (LIS) .
Product
AutoMate 1250, Sorter with Aliquot Module - UDI: 15099590364106; Model #ODL25125 Automate 2550, High Speed Sorter with Aliquot Module - UDI: 15099590364083 , Model #ODL25255
Manufacturer(s)
Beckman Coulter Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →