Beckman Coulter Inc. recalls Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2

Recalled July 31, 2024 · FDA recall Z-2428-2024 · Beckman Coulter

Hazard
LIH Influence check settings for Lipemia are incorrect which may result patient samples being reported as false low results or cause a high result to report as normal · FDA Class II
Units
13,565 units
Sold at
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, WA, WI, WV, WY and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China (DX), Colombia, Croatia, Cyprus, Czechia, Denmark, Ecuador, Egypt, Estonia, France, Georgia, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Korea, Republic of, Lebanon, Lithuania, Malaysia, Mayotte, Mexico, Morocco, Mozambique, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Province of China, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, United States of America, Viet Nam, Yemen.

Description

LIH Influence check settings for Lipemia are incorrect which may result patient samples being reported as false low results or cause a high result to report as normal

Product
Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2
Manufacturer(s)
Beckman Coulter Inc.