Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137

Recalled March 19, 2025 · FDA recall Z-1326-2025 · Beckman Coulter

Hazard
The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results · FDA Class II
Units
245 units
Sold at
US Nationwide. Global Distribution.

Description

The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results

Product
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137
Manufacturer(s)
Beckman Coulter, Inc.