Home / Recalls / Medical Devices
Beckman Coulter Inc. recalls Lactate Reagent (LACT 2 x 50), Catalog No. A95550 Intended for the quantitative determination of Lactate concentratio
Recalled May 24, 2017 · FDA recall Z-2078-2017 · Beckman Coulter
Hazard
Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results. · FDA Class II
Units
92,837 units total (71,271 units in US)
Sold at
Worldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Netherlands, New Zealand, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States including Puerto Rico, and Vietnam.
Description
Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.
Product
Lactate Reagent (LACT 2 x 50), Catalog No. A95550 Intended for the quantitative determination of Lactate concentration in human plasma and cerebrospinal fluid (CSF).
Manufacturer(s)
Beckman Coulter Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →