Home / Recalls / Medical Devices
Beckman Coulter Inc. recalls Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnosti
Recalled September 20, 2017 · FDA recall Z-3118-2017 · Beckman Coulter
Hazard
Reagent carryover from Sekisui ACTH when utilized on the BEC AU 680 analyzer may lead to falsely elevated BEC Lipase recovery exceeding the total precision specification. · FDA Class II
Units
320 units
Sold at
Distribution US nationwide.
Description
Reagent carryover from Sekisui ACTH when utilized on the BEC AU 680 analyzer may lead to falsely elevated BEC Lipase recovery exceeding the total precision specification.
Product
Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnostics (UK) Limited, United Kingdom, and Sekisui Diagnostics P.E.I. Inc., Canada. Used in laboratory testing.
Manufacturer(s)
Beckman Coulter Inc.