Beckman Coulter Inc. recalls Triglycerides Reagent (TG 2 x 300), Catalog No. 445850. Intended for quantitative determination of triglycerides conc

Recalled May 24, 2017 · FDA recall Z-2083-2017 · Beckman Coulter

Hazard
Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results. · FDA Class II
Units
124,068 units total (47,920 units in US)
Sold at
Worldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Netherlands, New Zealand, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States including Puerto Rico, and Vietnam.

Description

Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.

Product
Triglycerides Reagent (TG 2 x 300), Catalog No. 445850. Intended for quantitative determination of triglycerides concentration in human serum or plasma.
Manufacturer(s)
Beckman Coulter Inc.