RecallWatchUSA

Home / Recalls / Medical Devices

Beckman Coulter Inc. recalls Uric Acid Reagent (URIC 2 x 300), Catalog No. 442785. Intended for the quantitative determination of uric acid concen

Recalled May 24, 2017 · FDA recall Z-2079-2017 · Beckman Coulter

Hazard
Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results. · FDA Class II
Units
100,286 units total (20,732 units in US)
Sold at
Worldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Netherlands, New Zealand, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States including Puerto Rico, and Vietnam.

Description

Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.

Product
Uric Acid Reagent (URIC 2 x 300), Catalog No. 442785. Intended for the quantitative determination of uric acid concentration in human serum, plasma or urine.
Manufacturer(s)
Beckman Coulter Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Beckman Coulter Inc. recalls Uric Acid Reagent (URIC 2 x 300), Catalog No. 442785. Intended for the quantitative determination of uric acid concen | RecallWatchUSA