Home / Recalls / Medical Devices
Beckman Coulter Inc recalls Uric Acid Reagent (URIC 2 X 300), REF: 442785, For In Vitro Diagnostic Use, Rx Only, UDI: 15099590575281
Recalled March 11, 2020 · FDA recall Z-1424-2020 · Beckman Coulter
Hazard
It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment. · FDA Class II
Units
16,983 reagents
Sold at
Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine,Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming, Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas,Utah,Virginia,Washington,West Virginia, Wisconsin,Wyoming Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Curacao, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Guyana, Hong Kong, India, Ireland, Italy, Japan, Kuwait, Lebanon, Lithuania, Malaysia, Mayotte, Mexico, Netherlands, Oman, Panama, Paraguay, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Rwanda, San Marino, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Swaziland, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, United States, Vietnam
Description
It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.
Product
Uric Acid Reagent (URIC 2 X 300), REF: 442785, For In Vitro Diagnostic Use, Rx Only, UDI: 15099590575281
Manufacturer(s)
Beckman Coulter Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →