RecallWatchUSA

Home / Recalls / Medical Devices

Cardiovascular Systems, Inc. recalls Diamondback 360 Coronary Orbital Atherectomy System, Model Number: DBEC-125. The system is indicated to facilitate st

Recalled April 1, 2015 · FDA recall Z-1290-2015 · Cardiovascular Systems

Hazard
The OAD saline sheath outer layer may flake off from the catheter due to friction between the outer material layer and corresponding materials it may come in contact with during clinical use. If the material is large enough, it has the potential to cause a flow limiting embolism. · FDA Class I
Units
30
Sold at
Nationwide Distribution-including the states of AZ, FL, GA, LA, MA, MS, MO, NJ, NY, OH, and RI.

Description

The OAD saline sheath outer layer may flake off from the catheter due to friction between the outer material layer and corresponding materials it may come in contact with during clinical use. If the material is large enough, it has the potential to cause a flow limiting embolism.

Product
Diamondback 360 Coronary Orbital Atherectomy System, Model Number: DBEC-125. The system is indicated to facilitate stent delivery in patients with coronary artery disease (CAD) who are acceptable candidates for PTCA or stenting due to de novo, severely calcified coronary artery lesions.
Manufacturer(s)
Cardiovascular Systems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov