December 29, 2021Medical DevicesFDA
Cardiovascular Systems Inc recalls WIRION EMBOLIC PROTECTION SYSTEM, Delivery Catheter and Filter and Retrieval Catheter, 6Fr, REF (Model Number) WRN-D6
Hazard: Under certain circumstances, the WIRION filter assembly may become difficult to withdraw. Under such circumstances, alternative means may be necessary for withdrawal and the device may be prone to tearing or separating which may result in a risk for embolization.
September 8, 2021Medical DevicesFDA
Cardiovascular Systems Inc recalls STEALTH 360 GEN2 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM, 1.25MM MICRO, 145 CM, REF PRD-SC30-MICRO
Hazard: There is a potential mislabeling of the OAD crown sizes: 1.50mm Sold Crown OADs may be packaged s 1.25mm Micro Crown OADs and vice versa.
September 8, 2021Medical DevicesFDA
Cardiovascular Systems Inc recalls DIAMONDBACK 360 GEN2 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM, 1.5MM SOLID, 145 CM, REF DBP-150SOLID145
Hazard: There is a potential mislabeling of the OAD crown sizes: 1.50mm Sold Crown OADs may be packaged s 1.25mm Micro Crown OADs and vice versa.
September 19, 2018Medical DevicesFDA
Cardiovascular Systems Inc recalls CSI, Orbital Atherectomy System (OAS) Saline Infusion Pump, Rx only, Part Number: 7-10014, Model Number: SIP-3000. P
Hazard: The pumps may switch to stand-by during use requiring the pump to be reset prior to continuing treatment.
February 14, 2018Medical DevicesFDA
Cardiovascular Systems Inc recalls Peripheral Diamondback 1.50 Solid OAD, a percutaneous orbital atherectomy system indicated for use as therapy in patient
Hazard: Cardiovascular Systems, Inc. (CSI) is removing the products because it was identified that the labeling for two lots of product s were potentially inadvertently switched during production.
February 14, 2018Medical DevicesFDA
Cardiovascular Systems Inc recalls Peripheral Diamondback 1.25 Solid OAD, a percutaneous orbital atherectomy system indicated for use as therapy in patient
Hazard: Cardiovascular Systems, Inc. (CSI) is removing the products because it was identified that the labeling for two lots of product s were potentially inadvertently switched during production.
May 31, 2017Medical DevicesFDA
Cardiovascular Systems Inc recalls CSI, Orbital Atherectomy System (OAS) Saline Infusion Pump, Rx only, Part Number: 7-10014, Model Number: SIP-3000.
Hazard: Cardiovascular Systems, Inc. (CSI). has initiated a recall to remove 1,396 7-10014 Saline Infusion Pumps (SIP) which were distributed by CSI between 07 April 2015 and 04 April 2017. The 7-10014 pumps may switch to stand-by during use requiring the pump to be reset prior to continuing treatment. To date, none of the associated complaints have resulted in any patient harm.
May 3, 2017Medical DevicesFDA
Cardiovascular Systems Inc recalls ViperWire Advance with Flextip Peripheral Atherectomy Guide Wire, Part Number: 7-10026-01,Model Number: VPR-GW-FT14.
Hazard: CSI is recalling three lots of VIPERWIRE ADVANCE WITH FLEXTIP PERIPHERAL ATHERECTOMY GUIDEWIRE . It was discovered that 0.018in ViperWire Advance with Flextip product was packaged as a 0.014in ViperWire Advance with Flextip.
April 19, 2017Medical DevicesFDA
Cardiovascular Systems Inc recalls ViperWire Advance Peripheral Guide Wire with Flex Tip, Model No. VPR-GW-FT18. Sterilized with Ethylene Oxide. The
Hazard: Cardiovascular Systems, Inc. is recalling one lot of ViperWire Advance with Flextip, model VPR-GW-FT18 because it may contain a 0.014" ViperWire Advance Flextip product instead of the 0.018" ViperWire Advance with Flextip.
March 1, 2017Medical DevicesFDA
Cardiovascular Systems Inc recalls DIAMONDBACK 360 Peripheral 1.50 Classic, Model DBP-150CLASS145 Peripheral Orbital Atherectomy System (OAS) is a minimal
Hazard: CSI discovered that a 1.50mm Solid Crown OAD was labeled as a 1.50mm Crown OED.
July 6, 2016Medical DevicesFDA
Cardiovascular Systems Inc recalls ViperWire Advance Peripheral Guide Wire, Sterilized with Ethylene Oxide, Rx Only. Part number 72023-02, Model Number VPR
Hazard: The pouch label was missing the use by date (UBD) of 2018-04.
April 1, 2015Medical DevicesFDA
Cardiovascular Systems, Inc. recalls Diamondback 360 Coronary Orbital Atherectomy System, Model Number: DBEC-125. The system is indicated to facilitate st
Hazard: The OAD saline sheath outer layer may flake off from the catheter due to friction between the outer material layer and corresponding materials it may come in contact with during clinical use. If the material is large enough, it has the potential to cause a flow limiting embolism.
August 20, 2014Medical DevicesFDA
Cardiovascular Systems, Inc. recalls CSI Cardiovascular Systems, Inc., Diamondback 360 Peripheral Orbital Atherectomy System, Model Number DBP-125MICRO145, P
Hazard: CSI has initiated a recall on the Diamondback 360 Peripheral Orbital Atherectomy Device because it may contain a saline sheath that may experience cracking, fracture, and release particulate during use.
March 5, 2014Medical DevicesFDA
Cardiovascular Systems, Inc. recalls CSI, Diamondback 360 Coronary Orbital Atherectomy System, Part Number 70058-02, Model Number DBEC-125 Product Usage:
Hazard: Cardio vascular Systems Inc. has initiated a recall to request the immediate removal and return of Diamondback 360 Coronary Orbital Atherectomy Device (OAD). The device mistakenly contains a saline line that is used in our peripheral atherectomy device. This saline line was not part of the FDA approval for use in the coronary device. It therefore does not meet specification for this device. C
July 18, 2012Medical DevicesFDA
Cardiovascular Systems, Inc. recalls CSI Stealth 360-degree Orbital Atherectomy Device (OAD), Part Number 70056-11, Model Number PRD-SC30-125S. Sterilized w
Hazard: CSI has initiated a recall on specific lot numbers of Stealth 360-degree Orbital Atherectomy devices which have the potential for an insufficient crown solder bond that may result in the crown becoming detached from the driveshaft.