Cardiovascular Systems Inc recalls WIRION EMBOLIC PROTECTION SYSTEM, Delivery Catheter and Filter and Retrieval Catheter, 6Fr, REF (Model Number) WRN-D6

Recalled December 29, 2021 · FDA recall Z-0394-2022 · Cardiovascular Systems

Hazard
Under certain circumstances, the WIRION filter assembly may become difficult to withdraw. Under such circumstances, alternative means may be necessary for withdrawal and the device may be prone to tearing or separating which may result in a risk for embolization. · FDA Class I
Units
711 units
Sold at
United States- AK, AL, AR, AZ, CA, DC, FL, GA, GU, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MS, ND, NE, NH, NJ, NM, NV, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV OUS - Netherlands

Description

Under certain circumstances, the WIRION filter assembly may become difficult to withdraw. Under such circumstances, alternative means may be necessary for withdrawal and the device may be prone to tearing or separating which may result in a risk for embolization.

Product
WIRION EMBOLIC PROTECTION SYSTEM, Delivery Catheter and Filter and Retrieval Catheter, 6Fr, REF (Model Number) WRN-D6
Manufacturer(s)
Cardiovascular Systems Inc