Dako North America Inc. recalls PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor inte

Recalled November 21, 2012 · FDA recall Z-0287-2013 · Dako North America

Hazard
Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open. · FDA Class II
Units
136 units
Sold at
Nationwide distribution: USA including states of: AK, CA, CO, CT, DC, FL, GA, HI, IA, IL IN, KS, KY, MA, MD, MI, MN, MO, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, and WA.

Description

Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.

Product
PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.
Manufacturer(s)
Dako North America Inc.