July 18, 2018Medical DevicesFDA
Dako North America Inc. recalls PD-L1 IHC 22C3 pharmDx is a companion diagnostic used to determine PD-L1 levels for determination of patient treatment o
Hazard: To correct the kit's instructions for use as the storage time for gastric and gastroesophageal junction adenocarcinoma cut sections will be changed from six months to five months.
July 6, 2016Medical DevicesFDA
Dako North America Inc. recalls Dako HER2 CISH pharmDx Kit, product code: SK 109 Product Usage: For in vitro diagnostic use. HER2 CISH pharmDx Kit
Hazard: An incorrect HER2 probe concentration was used when producing the probe component (Vial 3) in this kit lot. This has resulted in the HER2 probe concentration being too low.
November 11, 2015Medical DevicesFDA
Dako North America Inc. recalls EnVision FLEX/HRP visualization reagent found in the following kits: EnVision FLEX, High pH (Link) EnVision FLEX+, Mou
Hazard: There is a defect in one lot of a buffer, non-specific background staining that has, in some instances, interfered with interpretation of the staining results.
May 6, 2015Medical DevicesFDA
Dako North America Inc. recalls Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated slide stai
Hazard: If a user requests slides from the LIS or TPID, then updates a request by changing the test, the TRD software will reject the update. The LIS and printed slide label will display the updated test with the requested change, but the Omnis system will execute the initial test, not reflecting the change. The system will not warn the user that the test request was rejected.
March 18, 2015Medical DevicesFDA
Dako North America Inc. recalls Autostainer Link 48 (AS480), Autostainer Plus Link Instrument (AS100), Autostainer (S3400), and Autostainer Plus (S3800)
Hazard: A false negative result affecting the diagnosis may occur. A defect in the syringe assembly stopcock component may, in some circumstances, cause leakage of excess buffer onto slide location 34 on the AS480, or slide locations 35 and 36 on the AS100, S3400, and S3800. Affected dates are from 12/2013-08/2014.
October 1, 2014Medical DevicesFDA
Dako North America Inc. recalls Dako Autostainer Link 48 with software version Dako Link 4.0.3, Model No. AS480 Product Usage: For in vitro diagnost
Hazard: Dako is recalling the Dako Link 4.0.3 software on the Dako Autostainer Link 48 because Dako has determined that an upgrade to this version of the software may cause premature rejection of reagent bottles or inadequate aspiration of reagent if the instrument has also had a preventative maintenance or waste pump replacement, and the baseplate is not completely aligned.
March 5, 2014Medical DevicesFDA
Dako North America Inc. recalls Product Name: HER2 CISH pharmDx Kit Catalog/Model number: SK109, lot 20000910 Product Usage: For in vitro diagnosti
Hazard: Dako has initiated a recall of HER2 CISH pharmDxTM Kit because weak red signals have been observed.
November 21, 2012Medical DevicesFDA
Dako North America Inc. recalls PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor inte
Hazard: Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.