Dako North America Inc. recalls Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated slide stai

Recalled May 6, 2015 · FDA recall Z-1548-2015 · Dako North America

Hazard
If a user requests slides from the LIS or TPID, then updates a request by changing the test, the TRD software will reject the update. The LIS and printed slide label will display the updated test with the requested change, but the Omnis system will execute the initial test, not reflecting the change. The system will not warn the user that the test request was rejected. · FDA Class II
Units
33 (7 in U.S. and 26 Internationally)
Sold at
Worldwide Distribution - U.S Nationwide in the states of IN and PA and the countries of: Spain, Belgium, France, Netherlands, Denmark, Finland, Sweden, Germany, UK, and Korea.

Description

If a user requests slides from the LIS or TPID, then updates a request by changing the test, the TRD software will reject the update. The LIS and printed slide label will display the updated test with the requested change, but the Omnis system will execute the initial test, not reflecting the change. The system will not warn the user that the test request was rejected.

Product
Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated slide stainer for in vitro diagnostic use. The TRD is intended to distribute patient case, appropriate parts information, track changes, and transform test requests from LIS or manual entries to connected Dako systems.
Manufacturer(s)
Dako North America Inc.