RecallWatchUSA

Home / Recalls / Medical Devices

Fort Defiance Industries, LLC recalls Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Recalled August 26, 2026 · FDA recall Z-2983-2026 · Fort Defiance Industries

Hazard
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path. · FDA Class II
Sold at
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.

Description

The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.

Product
Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
Manufacturer(s)
Fort Defiance Industries, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Fort Defiance Industries, LLC recalls Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131 | RecallWatchUSA