RecallWatchUSA

Home / Recalls / Medical Devices

Full Vision Inc recalls FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST1 110V, Model #317-07926GE

Recalled December 14, 2022 · FDA recall Z-0368-2023 · Full Vision

Hazard
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop. · FDA Class II
Units
69 units
Sold at
Wi, GA, NY, Mexico

Description

This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.

Product
FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST1 110V, Model #317-07926GE
Manufacturer(s)
Full Vision Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov