Home / Recalls / Medical Devices
Full Vision Inc recalls FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST2 220V, Model #317-07927GE
Recalled December 14, 2022 · FDA recall Z-0369-2023 · Full Vision
Hazard
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop. · FDA Class II
Units
165 units
Sold at
Wi, GA, NY, Mexico
Description
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
Product
FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST2 220V, Model #317-07927GE
Manufacturer(s)
Full Vision Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →