RecallWatchUSA

Home / Recalls / Medical Devices

Getinge Disinfection Ab recalls Washer Disinfector Aquadis 56; Model Number: 56A

Recalled May 14, 2025 · FDA recall Z-1720-2025 · Getinge Disinfection Ab

Hazard
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing. · FDA Class II
Units
102 units (8 US, 94 OUS)
Sold at
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Brazil, Canada, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Italy, Japan, Kenya, Korea, Kuwait, Latvia, Lithuania, Malaysia, Mauritania, Mexico, Netherlands, New Caledonia, Nigeria, Norway, Oman, Pakistan, Philippines, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Vietnam, Zambia.

Description

The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.

Product
Washer Disinfector Aquadis 56; Model Number: 56A
Manufacturer(s)
Getinge Disinfection Ab
Made in
Sweden

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Getinge Disinfection Ab recalls Washer Disinfector Aquadis 56; Model Number: 56A | RecallWatchUSA