October 8, 2025Medical DevicesFDA
Getinge Disinfection Ab recalls Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
Hazard: Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
May 14, 2025Medical DevicesFDA
Getinge Disinfection Ab recalls Washer Disinfector Aquadis 56; Model Number: 56M
Hazard: The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
May 14, 2025Medical DevicesFDA
Getinge Disinfection Ab recalls Washer Disinfector Aquadis 56; Model Number: 56A
Hazard: The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
April 25, 2018Medical DevicesFDA
Getinge Disinfection Ab recalls Getinge AGS, FSLC, FSUC, RC accessory to Washer- Disinfector 88 Series UDI Info: 07340153700109, 07340153700116 Ge
Hazard: There is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected goods could end up falling to the floor causing risk to nearby users and damage to goods and equipment.
April 25, 2018Medical DevicesFDA
Getinge Disinfection Ab recalls Getinge AGS, FSLC, FSUC, RC accessory to Washer- Disinfector 86 Series UDI Info: 86 Series Washer Disinfector 07340153
Hazard: There is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected goods could end up falling to the floor causing risk to nearby users and damage to goods and equipment.
August 26, 2015Medical DevicesFDA
Getinge Disinfection Ab recalls Getinge 86 Series Washer/Disinfector with G1 control system software, Model 8666 and Model 8668. For use in healthcare f
Hazard: Getinge Disinfection AB has initiated a Field Correction due to the potential for unauthenticated remote access to the product over the network.
December 31, 2014Medical DevicesFDA
Getinge Disinfection Ab recalls GETINGE 46-SERIES Medical Washer-Disinfector, Models 46-4 and 46-5, with PACS 300/350 Control Series with dryer --- Clas
Hazard: Getinge Disinfection AB received complaints regarding sediment residuals in the manifold of the washer and disinfector of Getinge 46-Series (CM302) with PACS 300/350 control. After performing detailed root cause analysis, the firm has identified the problem as insufficient water flow throughout the manifold due to formation of air pockets where sedimentation may occur.