Home / Recalls / Medical Devices
Getinge Disinfection Ab recalls Washer Disinfector Aquadis 56; Model Number: 56M
Recalled May 14, 2025 · FDA recall Z-1721-2025 · Getinge Disinfection Ab
Hazard
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing. · FDA Class II
Units
424 units (79 US, 345 OUS)
Sold at
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Brazil, Canada, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Italy, Japan, Kenya, Korea, Kuwait, Latvia, Lithuania, Malaysia, Mauritania, Mexico, Netherlands, New Caledonia, Nigeria, Norway, Oman, Pakistan, Philippines, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Vietnam, Zambia.
Description
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
Product
Washer Disinfector Aquadis 56; Model Number: 56M
Manufacturer(s)
Getinge Disinfection Ab
Made in
Sweden
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →