RecallWatchUSA

Home / Recalls / Medical Devices

Hamilton Medical AG recalls Hamilton-C3 Ventilator Unit Product Usage: The HAMILTON-C3 ventilator is intended to provide positive pressure venti

Recalled August 22, 2018 · FDA recall Z-2796-2018 · Hamilton Medical AG

Hazard
An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side. · FDA Class II
Units
2 devices
Sold at
US distribution to CA, CO, IL, LA, MN, NY, OK, VA, WI. Worldwide distribution to China, Germany, India, Indonesia, Israel, Kazakhstan, Korea, Pakistan, Taiwan, Poland, Armenia.

Description

An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.

Product
Hamilton-C3 Ventilator Unit Product Usage: The HAMILTON-C3 ventilator is intended to provide positive pressure ventilator support to adults, pediatrics, infants and neonates depending on its configuration. The device is used in the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room, and during transfer of ventilated patients within the hospital.
Manufacturer(s)
Hamilton Medical AG
Made in
Switzerland

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Hamilton Medical AG recalls Hamilton-C3 Ventilator Unit Product Usage: The HAMILTON-C3 ventilator is intended to provide positive pressure venti | RecallWatchUSA