Hamilton Medical AG recalls HAMILTON-G5, with software versions less than or equal to 2.60 The HAMILTON-G5/S1 ventilators are designed for intens

Recalled July 17, 2019 · FDA recall Z-1874-2019 · Hamilton Medical AG

Hazard
New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. When the ventilator enters the ambient state, alternative ventilation must be provided immediately. · FDA Class I
Units
4338 units
Sold at
Worldwide Distribution- US (Nationwide Distribution): MI, WI, OR, MO, TX, NV, PA, SC, HI, MN, ND, NC, AZ, FL, CA, MA, GA, NY, VA, WA, IN, CO, LA, IL, OH, TN, SD, OK, MI, AR, C, AK, WV, NJ, IA, NE, KS, MT, NH, AL, MD, KY, CT, and RI. And (International Distribution): AT, BE, BG, CH, CZ, DE, DK, EE, ES, FI, FR, GB, GR, HU, IT, MT, NL. NO, PL.PT, RO, SE, SI, SK, TR, HR. AGO, ARE, ARM, AUS, AZE, BGD, BHR , BLR, BOL, BRA, CAN, CHL, CHN, COL, CRI, DOM, DZA, ECU, EGY, EST, HKG, IDN, IND, IRQ, ISR, JOR, JPN, KAZ, KOR, KWT, LBN, LBY, LKA, LVA, MAR, MDA, MDV, MEX, MNG, MYS, NPL, NZL, OMN, PAK, PAN, PER, PHL, PRI, PSE, QAT, RUS, SAU, SGP, SUR, SYR, THA, TUN, TWN, UKR, URY, USA, VEN, VNM, YEM, and ZAF.

Description

New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. When the ventilator enters the ambient state, alternative ventilation must be provided immediately.

Product
HAMILTON-G5, with software versions less than or equal to 2.60 The HAMILTON-G5/S1 ventilators are designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The device is intended for use in the hospital and institutional environment where health care professionals provide patient care.
Manufacturer(s)
Hamilton Medical AG
Made in
Switzerland

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov