RecallWatchUSA

Home / Recalls / Medical Devices

Handicare Usa Inc recalls PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered

Recalled October 18, 2017 · FDA recall Z-0013-2018 · Handicare Usa

Hazard
During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop. · FDA Class II
Units
8 units
Sold at
Us and Canada

Description

During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.

Product
PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered
Manufacturer(s)
Handicare Usa Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Handicare Usa Inc recalls PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered | RecallWatchUSA