5 recalls on record · May 20, 2015 to August 29, 2018 · Medical Devices

Recalls by year

15
16
17
18

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

August 29, 2018Medical DevicesFDA
Handicare Usa Inc recalls Handicare P600 Series Patient Lift Product Usage: A device used to lift and transport a patient from one place to anot
Hazard: Premature strap wear and breakage at maximum weight conditions (625 lbs.).
August 29, 2018Medical DevicesFDA
Handicare Usa Inc recalls Handicare C Series Patient Lift Product Usage: A device used to lift and transport a patient from one place to anoth
Hazard: Premature strap wear and breakage at maximum weight conditions (625 lbs.).
October 18, 2017Medical DevicesFDA
Handicare Usa Inc recalls PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered
Hazard: During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.
November 2, 2016Medical DevicesFDA
Handicare USA, Inc. recalls ReTurn 7500/7500 Aid, transfer
Hazard: During the period of April 2009  May 2011, Handicare AB used a wing handle in which the screw and handle, after repeated tightening, may slide apart. This causes a small amount of play in the attachment of the ladder, which in turn results in high load at the attachment holes of the ladder. Eventually, this may cause mechanical fatigue in the material of the ladder.
May 20, 2015Medical DevicesFDA
Handicare USA, Inc. recalls ReTurn belt, wipeable, sizes Small through Extra Large, Plastic pouch, 1 item per pouch. Patient transfer aid.
Hazard: Complaints were received that the heat-welded seam on the wipeable belt's strap tore when force was applied.